This is the setup for phase contrast the Olympus BX and CH microscopes.
After the invention of the achromat lens the second most significant development in microscopy was phase contrast. This system allows the microscopist to "cheat" the classic optical tradeoff of resolution for contrast. So you are able to have resolution and contrast at the same time. In addition, it also allows for the examination of live specimens. Before phase contrast specimens had to be stained, with some dye that produces contrast by absorbing light. Since, the dye attaches itself to a particular biological molecule, it generally renders it lifeless. Most of these stains used in biology are toxic to the specimen they stain. The reason phase contrast is not toxic to the specimen is that it does not generate contrast by absorbing light, which is the way most contrast is formed. So there are no stains required. Phase contrast uses interference to generate contrast.
![]() |
![]() |
![]() |
Most phase contrast microscopes used today actually use two ways to produce non-absorption contrast. Interference contrast and stop contrast. In stop contrast an annulus slot is placed below the condenser lens. The cone of light that passes through the specimen is refracted slightly more in material that has a higher index of refraction. In the objective there is another annulus. If the specimen did not refract the light, it would be able to make it through the object annulus and ultimate to your eye. However if it was refracted it would not pass and the specimen would appear dark.
![]() |
1. Select the phase annuli turret in the condenser lens assembly to "O" for brightfield operation.
![]() |
2. Using a specimen with the appropriate slide and cover glass, set up the microscope for Kohler illumination. This must be done correctly or phase contrast will not work.
![]() | If you don't remember how to setup Kohlor illumination , click here. |
3. Select the either the 10X, 20X, 40X or 100X objective.
![]() |
4. Depending on which objective you are using, select the appropriate phase annulus from the table below.
| Objective | BX40 Phase Turret | BH Phase Turret |
| 4X | N/A | N/A |
| 10X | Ph1 | 10 |
| 20X | Ph1 | 20 |
| 40X | Ph2 | 40 |
| 100X oil | Ph3 | 100 |
5. Open the iris diaphragm all the way.
![]() |
6. Remove the eyepiece, and visually check the annulus and the phase plate are aligned. If not, use the annulus alignment knobs located on the condenser lens assembly to correct the alignment.
![]() |
7. Replace the eyepiece.
![]() |